YK-8000A:VITAL SIGNS MONITOR

product details

YK-8000AVITAL SIGNS MONITOR

TECHNICAL SPECIFICATIONS

Quality Standards

CE, ISO13485, ISO9001,

Design patent certification

 

Application Range

Adult/Pediatric/Neonatal/Medicine/Surgery/

Operating Room/ICU/CCU

 

Display

7” real color TFT screen

 

Environment

Operating environment :

Temperature: 0 ~ 40

Humidity: 15% ~ 80%

Pressure: 86.0 kPa ~ 106.0kPa

Transport and Storage environment :

Temperature: -20 ~ 55 ℃

Humidity: 10% ~ 90%

Pressure: 86.0 kPa ~ 106.0kPa

 

Power Requirements

AC: 100 ~ 240V, 50Hz/60Hz

DC: Built-in rechargeable battery

Battery: 11.1V 24wh Li-ion battery

2h operating after full charge

5min operating after low battery alarm

 

Dimension and Weight

Equipment:

175mm × 140mm ×300 mm; 3.3 kg

Package:

360mm ×250 mm × 300mm; 4.5 kg

 

ST segment

ST Segment Range: -2.0mV ~ +2.0mV

Accuracy: 0.02mV

 

SpO2

Measurement range: 0 ~ 100%

Resolution: 1%

Accuracy: 70% ~ 100% ± 2%

40% ~ 69% ± 3%

 

Pulse Rate

Measure/alarm range: 20 ~ 300bpm

Resolution: 1bpm

Accuracy: ± 3bpm

 

Displayed Parameters

Time: Battery-backed quartz crystal clock

ECG: ECG Waveform Scale, displayed lead

Heart Rate: Derived from ECG or SpO2

NIBP: Pressure (systolic, mean and diastolic)

Pulse Oximeter: Pulse Rate, SpO2 waveform, and

Oxygen saturation.

Respiration Rate: Respiration rate derived from ECG.

Trends: HR, RR, NIBP, TEMP and SpO2

Temperatures: one channels

ECG

5 Leads: RA, LA, LL, RL, V

Lead mode: ,Ⅱ, Ⅲ, AVR, AVL, AVF, V

Gain: × 1.25mm/mV,2.5mm/mV,

5.0mm/mV, 10mm/mV,

20mm/mV (error ≤ 5%)

Sweep speed: 6.25mm/s,12.5mm/s,25mm/s,

50mm/s (error ≤ ± 10%)

Bandwidth: Surgery 1 ~ 25 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

Sensitivity: >200 μ V P-P

CMRR: ≥ 112dB

Differential Input Impedance: >5M

ohm Electrode Offset Potential: ±300mV

Polarization Voltage: ≥ ± 500mV

Input Impedance: ≥ 5 M

Calibration Voltage: 1mV ± 5%

RESP

Method: impedance between RA-LL

Resp Impedance range: 0.3 ~ 3 Ω

Base Impedance range: 200 Ω ~ 4000 Ω

Bandwidth: 0.3 ~ 2.5Hz

Gain: × 0.25, × 0.5, × 1, × 2, × 4

Resp rate:

Measurement Range

Adult 0 ~ 120BrPM

Neonatal/Pediatric 0 ~ 150BrPM

Resolution: 1BrPM

Accuracy: 7 ~ 150BrPM ( ± 2BrPM)

0 ~ 6BrPM (unspecified)

Apnea Alarm: 10 ~ 40s

 

NIBP

Method: Oscillometry

Measure mode: Manual, Auto, STAT

Measure interval in auto mode:

1~5/10/15/30/60/90/120/180/240/480min

Measure period in STAT mode: 5min

Pulse rate range: 40 ~ 240bpm

Measure & alarm range: Adult

SYS 40 ~ 280mmHg

DIA 10 ~ 220mmHg

MEAN 20 ~ 240mmHg

Pediatric

SYS 40 ~ 220mmHg

DIA 10 ~ 160mmHg

MEAN 20 ~ 170mmHg

Neonatal

SYS 40 ~ 135mmHg

DIA 10 ~ 100mmHg

MEAN 20 ~ 110mmHg

Static pressure measurement

Range: 0 ~ 300mmHg

Accuracy: ± 3mmHg

Resolution: 1mmHg Accuracy:

Max. Mean Error: ± 5mmHg

Max. Standard Deviation: 8mmHg

Overpressure protection:

Adult 300mmHg

Pediatric 240mmHg

Neonatal 150mmHg

 

TEMP(optional)

Channel 1

Measure & alarm range: 0 ~ 50

Resolution: 0.1 ℃

Accuracy (no sensor): ±0.1 ℃ (0~50 ℃)

 

Standard accessories

NIBP cuff & tube

ECG cable & electrodes

SpO2 sensor

Lithium-ion battery

Power cable

Operator’s manual

 

Optional accessories

TEMP probe (skin)

Trolley bracket

Hanging bracket

Monitor recorder

Vehicle-mounted power supply