IE8:VITAL SIGNS MONITOR

product details

IE8VITAL SIGNS MONITOR 

Description

NIBP

Method Oscillometry

Measure mode: Manual Auto, STAT

Measure interval in auto mode

1-5/10115/3060/90/120/180/240/480m

Measure period in STAT mode: 5min

Pulse rate range: 40- 240bpm

Measure &alarm range:

Adult

SYS 40-280mmHg

DIA 10~220mmHg

MEAN 20-240mmHg

Pediatric

SYS 40~220mmHg

DIA 10-160mmHg

MEAN 20- 170mmHg

Neonatal

 

SYS 40~135mmHg

DIA 10- 100mmHg

MEAN 20-110mmHg

Static pressure measurement

Range: 0 -300mmHg

Accuracy: ±3mmHg

Resolution: 1mmHg Accuracy

Max Mean Error: ± 5mmHg

Max Standard Deviation: 8mmHg

Overpressure protection

Adult 300mmHg

Pediatric 240mmHg

Neonatal 150mmHg

 

Spo2

Measurement range: 0-100%

Resolution

Accuracy: 70%~100% ±2%

40%~69% ±3%

Environment

Operating environment:

Temperature: 0-40 C

Humidity: 15%-80%

Pressure: 86.0 kPa- 106.0kPa

Transport and Storage environment

Temperature:-20-55 C

10%~90%

Pressure: 86.0 kPa-106.0kPa

 

Quality Standards and Classification

CE, ISO13485

SFDA: Class b

Anti-electroshock degree:

Classequipment

(internal power supply)

TEMP/SpO 2 /NIBP: BF

ECG/Resp: CF

________________________________________________________________________________________________________________________________________________________________________________________________________

282-removebg-preview.png284-removebg-preview.png

286-removebg-preview.png生命体征监护仪IE8PNG-removebg-preview.png